Optimized RF Shimming for MRI in Patients with Deep Brain Stimulation Implants
Author Information
Author(s): Arianpouya Maryam, Yang Benson, Tam Fred, McElcheran Clare E., Graham Simon J.
Primary Institution: Physical Sciences Platform, Sunnybrook Research Institute, Toronto, ON, Canada
Hypothesis
The new workflow for RF shimming will yield improved B1+ mapping and reduced localized heating during MRI for patients with DBS implants.
Conclusion
The proposed workflow demonstrates improved B1+ uniformity and reduced localized heating compared to existing methods.
Supporting Evidence
- The proposed workflow produced smaller temperature elevations and improved signal uniformity compared to existing methods.
- Experimental results showed effective suppression of localized heating at the wire tip.
- The new workflow enhances the robustness of existing methods for MRI of DBS patients.
Takeaway
This study shows a new way to take pictures of the brain using MRI that helps keep patients with brain implants safe from overheating.
Methodology
The study used electromagnetic simulations and experimental PTX MRI to test a two-step optimization algorithm for B1+ mapping and RF shimming.
Limitations
The study is a proof of concept and does not establish a clinical workflow for MRI with DBS implants.