Reliability of Measuring First Metatarsophalangeal Joint Dorsiflexion
Author Information
Author(s): Curran Sarah, Jones Angela
Primary Institution: University of Wales Institute, Cardiff
Hypothesis
The study aims to determine the intra and interrater reliability of visual estimation and goniometry for assessing first metatarsophalangeal joint dorsiflexion.
Conclusion
Goniometry is more reliable than visual estimation for assessing first metatarsophalangeal joint dorsiflexion when performed by experienced examiners.
Supporting Evidence
- Experienced examiners showed excellent reliability for goniometry with ICC >0.953.
- Inexperienced examiners had fair-to-good reliability for goniometry with ICC 0.322 - 0.597.
- For visual estimation, inexperienced examiners had ICC 0.666 – 0.808.
- Goniometry's standard error of measurement was 1.8° – 2.5° for experienced examiners.
Takeaway
This study looked at how reliable two methods are for measuring a joint's movement, finding that one method is better when used by experienced people.
Methodology
Ten experienced and ten inexperienced examiners evaluated three clinical photographs on two separate occasions.
Limitations
Neither group's standard error of measurement was clinically acceptable.
Participant Demographics
Ten experienced and ten inexperienced examiners.