Pacemaker Prevention Therapy in Drug-refractory Paroxysmal Atrial Fibrillation: Reliability of Diagnostics and Effectiveness of Prevention Pacing Therapy in Vitatron™ Selection® Device
2006

Pacemaker Therapy for Atrial Fibrillation

Sample size: 15 publication Evidence: moderate

Author Information

Author(s): Terranova Paolo MD, Valli Paolo MD, Terranova Peppino MD, Dell’Orto Simonetta MD, Greco Enrico Maria MD

Primary Institution: University of Milan, Italy

Hypothesis

Can the Vitatron™ Selection® device effectively prevent atrial fibrillation in patients with brady-tachy syndrome?

Conclusion

The Vitatron™ device reliably detects atrial arrhythmias and significantly reduces their occurrence in patients with brady-tachy syndrome.

Supporting Evidence

  • 59 atrial arrhythmia episodes were recorded with a correlation coefficient of 0.96 between Holter and pacemaker data.
  • PACs decreased from 83±12/day to 2.3±0.8/day.
  • AF episodes reduced from 46±7/day to 0.12±0.03/day.
  • AF burden decreased from 93%±6% to 0.3%±0.06%.

Takeaway

This study shows that a special pacemaker can help people with a heart rhythm problem by stopping their heart from racing too much.

Methodology

15 patients with a DDDRP pacemaker were monitored for atrial arrhythmias and their effectiveness was evaluated over a follow-up period of 24±8 months.

Potential Biases

Potential selection bias due to the small number of patients and their specific characteristics.

Limitations

The study had a small and highly selected sample size, limiting generalizability.

Participant Demographics

9 males and 6 females, mean age of 71±5 years, NYHA class I-II.

Statistical Information

P-Value

p<0.001

Confidence Interval

95% confidence interval: 41.69 to 50.07

Statistical Significance

p<0.001