Radiographic viewing conditions at Johannesburg Hospital
2008

Radiographic Viewing Conditions at Johannesburg Hospital

Sample size: 47 publication Evidence: moderate

Author Information

Author(s): Thulani Nyathi, AN Mwale, P Segone, SH Mhlanga, ML Pule

Primary Institution: University of the Witwatersrand

Hypothesis

The study aims to measure the luminance level of X-ray viewing boxes and ambient lighting levels in reporting rooms as a quality assurance procedure.

Conclusion

The radiograph viewing conditions in Johannesburg variably comply with guidelines.

Supporting Evidence

  • The mean average luminance was 1026.75 cd m-2 at the Division of Radiology and 3284.38 cd m-2 at the Division of Radiation Oncology.
  • The average ambient lighting was found to be 66.30 lux and 66.43 lux at the Division of Radiation Oncology and Division of Radiology respectively.
  • The Division of Radiation Oncology had an average viewing box uniformity of 7.14% compared to 27.32% at the Division of Radiology.

Takeaway

This study looked at how well the lighting conditions for viewing X-rays at a hospital meet recommended standards, and found that they don't always match up.

Methodology

Luminance levels of X-ray viewing boxes and ambient lighting were measured using a calibrated photometer in reporting rooms.

Limitations

The study did not analyze compliance with European guidelines due to differences in measurement distance.