New Patellar Tendon Bearing Brace Study
Author Information
Author(s): Vahid Chamani, Mahmood Bahramizadeh, Mobina Khosravi, Akbar Biglarian, Gholamreza Ghorbani Amjad, Seyyed Mohammad Ebrahim Mousavi, Mokhtar Arazpour
Primary Institution: University of Social Welfare and Rehabilitation Sciences
Hypothesis
This study aimed to explore the relationship between a rise in displacement between the foot plate and calf shells of a new PTB brace and changes in the amount of offloading on the tibia.
Conclusion
Maintaining a distance of 1.5 cm or more between the calf shells and foot plate is effective in reducing weight and offloading on the tibia.
Supporting Evidence
- The new PTB brace provided varying levels of offloading, ranging from 22% to 38%.
- As the vertical distance between the calf shells and foot plate increased, the offloading levels also increased significantly.
- Significant differences in weight reduction and offloading percentages were observed when comparing 1 cm and 1.5 cm displacement.
Takeaway
The new brace helps reduce pressure on the leg by adjusting how far apart its parts are, making it better for healing broken bones.
Methodology
This pilot study tested a new PTB brace on five individuals with midshaft tibia fractures under six different conditions of displacement.
Limitations
The small sample size may limit the generalizability of the findings.
Participant Demographics
Participants had an average age of 45.8 years, height of 1.63 m, and body mass index of 27.32 kg/m2.
Statistical Information
P-Value
p=0.020
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website