Comparison of a new microplate oestrogen receptor (ER) enzyme immunoassay with other ER detection methods
1997

Comparison of a New Oestrogen Receptor Detection Method

Sample size: 50 publication Evidence: moderate

Author Information

Author(s): V Delagel, S Deytieux, V Le Doussal, F Degorcel, L Bellanger, K Hacene, P Seguin, F Descotes, S Saez, F Spyratos

Primary Institution: CIS Bio International

Hypothesis

How does the new microplate immunoassay for oestrogen receptors compare to existing detection methods?

Conclusion

The new microplate immunoassay (EIA96) is more sensitive than the radioligand assay while maintaining substantial agreement with other methods.

Supporting Evidence

  • EIA96 showed a sensitivity of 0.94 compared to 0.88 for RLA.
  • The positivity-negativity concordance of EIA96 with RT-PCR was 86%.
  • EIA96 and ER-EIA had the best correlation among the methods tested.
  • The area under the ROC curve for EIA96 was 0.881.

Takeaway

Researchers created a new test to find oestrogen receptors in breast cancer and found it works better than some older tests.

Methodology

The study compared five different methods for detecting oestrogen receptors in breast cancer tissue samples.

Limitations

The study involved a limited number of samples, which may affect the generalizability of the cut-off values.

Statistical Information

P-Value

2.4 x 10-7

Confidence Interval

[-3.170; -0.495]

Statistical Significance

p<0.05