Evaluation of the SF-3000 haematology analyser
1996

Evaluation of the SF-3000 Haematology Analyser

Sample size: 246 publication Evidence: moderate

Author Information

Author(s): Izumi Tsuda, Hitoshi Sagawa, Takayuki Takubo, Seiki Kawai, Noriyuki Tatsumi

Primary Institution: Osaka City University Medical School

Hypothesis

The SF-3000 haematology analyser will provide accurate and reliable measurements comparable to existing methods.

Conclusion

The SF-3000 haematology analyser demonstrated good precision and reproducibility, making it suitable for smaller laboratories.

Supporting Evidence

  • The SF-3000 showed good correlation with the SE-9000 for red cell and platelet parameters.
  • Reproducibility of the SF-3000 was confirmed with a mean CV for complete blood counts ranging from 0.3 to 2.8.
  • Linearity was observed in white blood cell, red blood cell, and platelet counts across various dilutions.

Takeaway

The SF-3000 is a small and affordable blood testing machine that works well for counting different types of blood cells.

Methodology

The study involved comparing the SF-3000 analyser with the SE-9000 analyser using blood samples processed within 6 hours.

Limitations

The study may not account for all variations in blood samples from different populations.

Participant Demographics

Technologists with more than 10 years of experience performed the tests.