An evaluation of the Technicon DPA*I specific protein analyser
1990

Evaluation of the Technicon DPA*I Protein Analyser

Sample size: 60 publication Evidence: moderate

Author Information

Author(s): M. J. Waterson, L. Skingle, C. P. Price

Primary Institution: Department of Clinical Biochemistry, The London Hospital

Hypothesis

The DPA*I analyser's performance in terms of precision, accuracy, and calibration stability will be evaluated across various analytes.

Conclusion

The DPA*I analyser demonstrated excellent precision and calibration stability, making it a reliable tool for protein analysis.

Supporting Evidence

  • The DPA*I analyser showed within-batch precision generally between 1% and 3%.
  • Calibration stability was maintained for a minimum of one month.
  • The instrument can perform approximately 100 tests per hour.

Takeaway

The DPA*I analyser is a machine that helps doctors quickly and accurately measure proteins in blood samples.

Methodology

The study assessed the DPA*I analyser's performance by analyzing various serum pools and comparing results with established methodologies.

Limitations

The study's precision data may have been affected by a short period of poor performance due to a blocked probe.