Laboratory evaluation of the COBAS MIRA S random access analyser
1991

Evaluation of the COBAS MIRA S Random Access Analyser

Sample size: 120 publication Evidence: moderate

Author Information

Author(s): Taweesook Kanluan, Srisanit Intaramanee, Surapon Tangvorasittichai, Orathai Tangvorasittichai, Lerson Suwanton, Sudarat Manochiopinij, Soontharee Tantrarongroj, Pimpan Leangphibul

Primary Institution: Mahidol University

Hypothesis

The study aims to assess the analytical performance of the COBAS MIRA S random access analyser in clinical laboratories.

Conclusion

The COBAS MIRA S is a reliable and efficient analyser for clinical chemistry tests, showing good correlation with the Hitachi 705.

Supporting Evidence

  • The COBAS MIRA S showed within-run precision values of % CV ranging from 0.54 to 3.37.
  • Correlation coefficients between the COBAS MIRA S and Hitachi 705 ranged from 0.95 to 0.99.
  • The instrument required minimal operator attention during operation.
  • Linearity testing covered the necessary pathological range for all analytes.

Takeaway

The COBAS MIRA S is a machine that helps test blood samples quickly and accurately, making it easier for doctors to get results.

Methodology

The study involved precision testing, linearity studies, and comparison with the Hitachi 705 using various clinical chemistry tests.

Limitations

The study was limited to a two-month evaluation period and specific analytes.

Participant Demographics

The study included more than 120 patient samples representing low, normal, and above-normal levels.

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication