Estudio comparativo de un nuevo dispositivo para el análisis de la calidad seminal y la evaluación microscópica manual en un laboratorio clínico no especializado
2024

Comparative Study of a New Device for Semen Quality Analysis

Sample size: 50 publication Evidence: moderate

Author Information

Author(s): Jani Erika, Bozzola Margherita, Zagler Elmar Marco, Roccaforte Vincenzo, Massimo Daves

Primary Institution: Laboratorio de Bioquímica Clínica, Hospital Provincial de Bolzano (SABES-ASDAA), Bolzano, Italia

Hypothesis

The study aims to evaluate the effectiveness of an automatic instrument for semen analysis compared to manual microscopy.

Conclusion

The LensHooke™ device shows an acceptable level of agreement with manual semen analysis.

Supporting Evidence

  • The median sperm concentration was higher in manual analysis compared to the LensHooke device.
  • The study found a moderate level of agreement in morphology and a very good agreement in motility between the two methods.
  • The Wilcoxon test showed no statistically significant differences between the two methods.

Takeaway

This study tested a new machine that checks sperm quality and found it works pretty well compared to the old way of looking at sperm under a microscope.

Methodology

50 semen samples were analyzed using both manual and automated methods simultaneously by experienced analysts.

Potential Biases

Subjectivity in manual analysis and reliance on analyst experience may introduce bias.

Limitations

The study had a small sample size and did not use the recommended staining for morphology evaluation.

Participant Demographics

Participants were aged 18-59 years.

Statistical Information

P-Value

p<0.05

Confidence Interval

95% CI for manual median: 27.60 to 75; 95% CI for LensHooke median: 19.9 to 66.27

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1515/almed-2024-0182