Contamination of RT-PCR Kits with Murine Leukemia Virus
Author Information
Author(s): Sato Eiji, Furuta Rika A, Miyazawa Takayuki
Primary Institution: Institute for Virus Research, Kyoto University
Hypothesis
Is the commercial RT-PCR kit contaminated with endogenous murine leukemia virus genomes?
Conclusion
The study found that a commercial RT-PCR kit was contaminated with RNA from a murine leukemia virus, which could lead to false-positive results in XMRV studies.
Supporting Evidence
- The RT-PCR kit from Invitrogen was found to be contaminated with RNA from a polytropic endogenous MLV.
- Positive results were consistently detected in negative controls using the contaminated kit.
- Contaminants were identified as RNA, as positive bands disappeared after adding RNaseA.
Takeaway
Some test kits used to find viruses might have tiny bits of other viruses in them, which can make it look like someone is sick when they aren't.
Methodology
The study evaluated four commercial one-step RT-PCR kits for contamination by amplifying specific viral gene regions and comparing sequences.
Potential Biases
Potential bias due to the specific kits tested and the possibility of contamination in laboratory settings.
Limitations
The study was limited to the evaluation of specific RT-PCR kits and may not represent all kits available on the market.