Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children
2006

Performance of a new hand-held device for measuring exhaled nitric oxide

Sample size: 71 publication Evidence: moderate

Author Information

Author(s): Alving K, Janson C, Nordvall L

Primary Institution: Karolinska Institutet

Hypothesis

The new hand-held device for measuring exhaled nitric oxide will perform comparably to the standard stationary device.

Conclusion

The hand-held device (NIOX MINO) and the stationary system (NIOX) are in clinically acceptable agreement, allowing for successful use in both adults and children.

Supporting Evidence

  • The hand-held device showed a high success rate of ≥ 84% for both adults and children.
  • The median of the intrasubject difference in exhaled NO for the two devices was -1.2 ppb, indicating slight higher readings from the hand-held device.
  • The study included both healthy controls and atopic patients with and without asthma.

Takeaway

Researchers tested a new small device that measures a gas called nitric oxide in breath, and it worked just as well as a big machine used in hospitals.

Methodology

The study compared a new hand-held device with a standard stationary device by measuring exhaled nitric oxide in 71 subjects, including both healthy individuals and asthma patients.

Limitations

Some younger children had difficulty using the hand-held device, and a few subjects were excluded due to low nitric oxide levels.

Participant Demographics

Participants included 36 males and a mix of healthy controls and atopic patients aged 6 to 60 years.

Statistical Information

P-Value

p<0.05

Confidence Interval

95% limits of agreement were -9.8 and 8.0 ppb.

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1465-9921-7-67