The reliability of toe systolic pressure and the toe brachial index in patients with diabetes
2010

Reliability of Toe Systolic Pressure and Toe Brachial Index in Diabetic Patients

Sample size: 30 publication Evidence: moderate

Author Information

Author(s): Romanos Mary T, Raspovic Anita, Perrin Byron M

Primary Institution: La Trobe University

Hypothesis

This study aimed to determine the intra and inter-rater reliability of the measurement of Toe Systolic Pressure and the Toe Brachial Index in patients with diabetes using a manual measurement system.

Conclusion

The study found that while toe systolic pressures and toe brachial indices have moderate to good reliability, the broad range of error raises concerns about their clinical utility.

Supporting Evidence

  • The mean intra-class correlation for intra-rater reliability for toe systolic pressures was 0.87.
  • The intra-class correlation for inter-rater reliability was 0.88 for toe systolic pressures.
  • Despite reasonable reliability, the broad range of error questions the clinical utility of the measurements.

Takeaway

Doctors measured blood flow in the toes of diabetic patients to see how reliable the measurements were, and they found that while the measurements can be useful, there's a lot of room for error.

Methodology

A repeated measures, reliability study where three raters measured Toe Systolic Pressure and Toe Brachial Index in thirty participants with diabetes on two occasions, one week apart.

Potential Biases

Potential sources of error include the equipment used, the experience of the raters, and the physiological status of the participants.

Limitations

The small sample size and the limited experience of raters may affect the reliability of the results.

Participant Demographics

The majority of participants were older males with type 2 diabetes, averaging 70 years of age and having diabetes for over 10 years.

Statistical Information

P-Value

p<0.01

Confidence Interval

95% CI: -25.97 to 26.06 mmHg for toe systolic pressure; -0.22 to 0.28 for toe brachial index.

Statistical Significance

p<0.01

Digital Object Identifier (DOI)

10.1186/1757-1146-3-31