The competency of the novel through‐the‐scope suture device for gastric mucosal defects: In vivo study in a porcine model (with video)
2024

Using a New Suture Device for Closing Stomach Wounds

Sample size: 24 publication Evidence: moderate

Author Information

Author(s): Ito Mamoru, Dobashi Akira, Tominaga Takanori, Futakuchi Toshiki, Tamai Naoto, Suka Machi, Sumiyama Kazuki

Primary Institution: The Jikei University School of Medicine

Hypothesis

Both expert and novice endoscopists can quickly become proficient in using the X‐Tack suture system to efficiently perform closures.

Conclusion

A small number of cases were required for both expert and novice endoscopists to reach competency in the novel through‐the‐scope suture device.

Supporting Evidence

  • Competency was achieved after six cases in the expert and seven cases in the novice.
  • The median closure time was 9.0 minutes for the expert and 8.0 minutes for the novice.
  • Complete closure success rates were 75.0% for the expert and 83.3% for the novice.
  • No adverse events were reported during the study.

Takeaway

Doctors can learn to use a new tool to stitch up stomach wounds quickly, even if they haven't used it before.

Methodology

Two endoscopists, one novice and one expert, closed 24 gastric mucosal defects in four pigs using the X‐Tack suture system.

Potential Biases

Both endoscopists had no prior experience with the device, which may affect the learning curve.

Limitations

The small sample size of endoscopists and limited number of defects may limit the generalizability of the findings.

Participant Demographics

Two endoscopists: one novice with less than 400 experiences and one expert with over 500 experiences.

Statistical Information

P-Value

0.590

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1002/deo2.70037