Low-Power, High Data Rate Transceiver System for Implantable Prostheses
Author Information
Author(s): A. R. Kahn, E. Y. Chow, O. Abdel-Latief, P. P. Irazoqui
Primary Institution: Weldon School of Biomedical Engineering, Purdue University
Hypothesis
The study investigates the effects of interferers and tissue attenuation on a wireless link for optimal design of future implantable systems.
Conclusion
The external device can demodulate data at 8 Mbps with a bit-error-rate better than 1e-5, making it suitable for high data-rate neural prostheses.
Supporting Evidence
- The transceiver achieved data rates of up to 8 Mbps with a bit-error-rate of 7.5e-5.
- Testing was conducted in various environments including an anechoic chamber and in vivo.
- The study highlights the importance of low power consumption in implantable devices.
Takeaway
This study shows how a special wireless system can send a lot of data from a tiny implant in your body without using much power.
Methodology
The study involved designing a custom transceiver and testing its performance in both controlled environments and in vivo.
Limitations
The study's results may not fully translate to human applications due to differences in tissue characteristics.
Participant Demographics
In vivo testing was performed on a rat model.