A novel miniaturized potentiometric electrode based on carbon nanotubes and molecularly imprinted polymer for the determination of lidocaine
2024

Miniaturized Electrode for Measuring Lidocaine

publication Evidence: high

Author Information

Author(s): Hassan Saad S. M., Fathy Mahmoud Abdelwahab

Primary Institution: Ain Shams University

Hypothesis

The integration of carbon nanotubes and molecularly imprinted polymers can enhance the sensitivity and selectivity of potentiometric electrodes for lidocaine detection.

Conclusion

The developed electrode demonstrated high sensitivity, fast response times, and excellent selectivity for lidocaine in various samples.

Supporting Evidence

  • The electrode showed a Nernstian slope of 58.92 ± 0.98 mV/decade.
  • It had a detection limit of 7.75 × 10−8 mol/l for lidocaine.
  • The electrode was successfully used for determining lidocaine in pharmaceutical preparations and human urine.

Takeaway

Researchers created a tiny sensor that can quickly and accurately measure lidocaine, a common anesthetic, using special materials that help it work better.

Methodology

The study involved developing a potentiometric screen-printed electrode using single-walled carbon nanotubes and a molecularly imprinted polymer for lidocaine detection.

Limitations

The study does not specify the long-term stability of the electrode in real-world applications.

Digital Object Identifier (DOI)

10.1007/s00604-024-06802-6