Blood transfer devices for malaria rapid diagnostic tests: evaluation of accuracy, safety and ease of use
2011

Evaluating Blood Transfer Devices for Malaria Tests

Sample size: 227 publication null Evidence: high

Author Information

Author(s): Heidi Hopkins, Wellington Oyibo, Jennifer Luchavez, Mary Lorraine Mationg, Caroline Asiimwe, Audrey Albertini, Iveth J González, Michelle L Gatton, David Bell

Primary Institution: Foundation for Innovative New Diagnostics

Hypothesis

Which blood transfer device is the most accurate, safe, and easy to use for malaria rapid diagnostic tests?

Conclusion

The inverted cup design achieved the highest overall performance for blood transfer devices used with malaria rapid diagnostic tests.

Supporting Evidence

  • The inverted cup was rated easiest to use by 91% of participants.
  • The calibrated pipette had the highest number of blood exposures at 10%.
  • Overall, the inverted cup was preferred by 72% of health workers.

Takeaway

This study looked at different tools for transferring blood for malaria tests and found that the inverted cup was the best one to use because it was easy and safe.

Methodology

Five blood transfer devices were tested by 227 health workers in Nigeria, the Philippines, and Uganda to evaluate their accuracy, safety, and ease of use.

Potential Biases

Participants were unaware of the provenance of the devices to avoid bias.

Limitations

The study used filter paper instead of actual RDT pads, which may have affected the results, especially for the glass capillary.

Participant Demographics

Participants included health workers from Nigeria, the Philippines, and Uganda, with varying levels of prior experience with blood transfer devices.

Statistical Information

P-Value

p<0.001

Confidence Interval

null

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.1186/1475-2875-10-30