Verifying 4D Gated Radiotherapy with Electronic Portal Imaging
Author Information
Author(s): van Sörnsen de Koste John R, Cuijpers Johan P, de Geest Frank GM, Lagerwaard Frank J, Slotman Ben J, Senan Suresh
Primary Institution: VU University medical center, Amsterdam, The Netherlands
Hypothesis
Can time-integrated electronic portal imaging (TI-EPI) effectively verify the position of internal structures during respiratory-gated radiotherapy (RGRT)?
Conclusion
TI-EPI is a promising method for verifying RGRT delivery, providing a direct link between anatomy and breathing patterns.
Supporting Evidence
- Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases.
- Cranio-caudal motion of internal structures ranged from 17.5–20.0 mm on planning 4DCT scans.
- Mean displacement in the two structures was 2.8 mm and 1.4 mm after bony alignment.
Takeaway
This study shows that a special imaging technique can help doctors make sure they are targeting the right part of the body during cancer treatment, even when the patient is breathing.
Methodology
The study used time-integrated electronic portal imaging (TI-EPI) to verify internal structure positions during RGRT, involving both phantom and patient data.
Potential Biases
Potential inter-observer variation in identifying anatomical structures was noted, although all registrations were performed by a single observer.
Limitations
The study's findings are based on a small sample size and may not be generalizable to all patients undergoing RGRT.
Participant Demographics
Two patients with lung and hepatic tumors undergoing RGRT.