Body composition in male elite athletes, comparison of bioelectrical impedance spectroscopy with dual energy X-ray absorptiometry
2008

Comparing Body Composition Methods in Male Athletes

Sample size: 33 publication Evidence: moderate

Author Information

Author(s): Svantesson Ulla, Zander Martina, Klingberg Sofia, Slinde Frode

Primary Institution: Sahlgrenska Academy, University of Gothenburg, Sweden

Hypothesis

How do body composition results from bioelectrical impedance spectroscopy compare with those from dual energy X-ray absorptiometry in male elite athletes?

Conclusion

Body composition results assessed by BIS in elite athletes should be interpreted with caution, especially in individual subjects.

Supporting Evidence

  • BIS underestimated fat mass by 4.6% points in ice hockey players.
  • BIS resulted in a lower mean fat mass by 1.1% points in soccer players.
  • Agreement between methods at the individual level was highly variable.

Takeaway

This study looked at how two different methods measure body fat in male athletes. It found that one method might not always give the right answer.

Methodology

Body composition was assessed using DXA and BIS at the same time, with agreement analyzed using paired t-tests.

Potential Biases

BIS showed low precision in estimating body composition at an individual level.

Limitations

Participants' conditions regarding exercise, dehydration, or fasting were not registered.

Participant Demographics

All participants were male elite athletes aged 18 years or older, competing in soccer and ice hockey.

Statistical Information

P-Value

0.0056

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1477-5751-7-1