Measurements of patient’s setup variation in intensity-modulated radiation therapy of head and neck cancer using electronic portal imaging device
2007

Measuring Patient Setup Variation in Radiation Therapy for Head and Neck Cancer

Sample size: 9 publication Evidence: moderate

Author Information

Author(s): Naiyanet N MSc, Oonsiri S MSc, Lertbutsayanukul C MD, Suriyapee S MEng

Primary Institution: King Chulalongkorn Memorial Hospital

Hypothesis

To measure the interfraction setup variation of patients undergoing intensity-modulated radiation therapy (IMRT) of head and neck cancer.

Conclusion

The population-based margin is less than 5 mm, thus the margin provides sufficient coverage for all of the patients.

Supporting Evidence

  • A total of 135 images and 405 match points were evaluated.
  • The systematic error ranged from 0 to 7.5 mm and the random error ranged from 0.3 to 4.8 mm.
  • The population-based margin ranged from 2.3 to 4.5 mm for the left-right direction.

Takeaway

This study looked at how much patients move during radiation therapy for head and neck cancer, and found that a smaller safety margin can still keep patients safe.

Methodology

Data was collected from 9 patients treated with dynamic IMRT, using electronic portal imaging to measure setup variations.

Limitations

The study is limited to a small sample size of 9 patients.

Participant Demographics

Patients were head and neck cancer patients treated at King Chulalongkorn Memorial Hospital.

Digital Object Identifier (DOI)

10.2349/biij.3.1.e30