Rationale for using insensitive quality control rules for today's hematology analyzers
2010

Quality Control in Hematology Analyzers

Sample size: 121 publication 10 minutes Evidence: moderate

Author Information

Author(s): George S. Cembrowski, B. Smith, D. Tung

Primary Institution: University of Alberta Hospital

Hypothesis

The study evaluates the stability of e-Check hematology quality control products in relation to the performance of Sysmex XE-2100 analyzers.

Conclusion

The recommended quality control limits for the Sysmex XE-2100 are acceptable and can be used effectively in laboratories.

Supporting Evidence

  • The study found significant variation in quality control practices across laboratories.
  • Most laboratories did not adhere to established quality control policies.
  • The recommended control rules can help reduce unnecessary re-analysis of samples.

Takeaway

This study looks at how well certain blood testing machines work and suggests better ways to check their accuracy.

Methodology

The study analyzed data from 121 laboratories using the Sysmex XE-2100 hematology analyzer over two 30-day periods.

Potential Biases

There may be biases in the reporting and adherence to quality control policies among participating laboratories.

Limitations

The study may not represent all hematology analyzers as it focuses on the Sysmex XE-2100.

Participant Demographics

Participants included hospital and non-hospital laboratories from primarily the US.

Statistical Information

P-Value

p<0.01

Statistical Significance

p<0.01

Digital Object Identifier (DOI)

10.1111/j.1751-553X.2010.01229.x