Performance characteristics of a novel blood bag in-line closure device and subsequent product quality assessment
2010

Evaluation of a New Blood Bag Closure Device

Sample size: 188 publication 10 minutes Evidence: moderate

Author Information

Author(s): Katherine Serrano, Elena Levin, Brankica Culibrk, Sandra Weiss, Ken Scammell, Wolfgang F. Boecker, Dana V. Devine

Primary Institution: Canadian Blood Services and The University of British Columbia Centre for Blood Research

Hypothesis

Does the new CompoFlow closure device affect the quality of blood components compared to standard closures?

Conclusion

The CompoFlow closure devices improved the level of process control and processing time of blood component production with no negative effects on product quality.

Supporting Evidence

  • The CompoFlow device significantly improved the mean extraction time.
  • PLT and RBC products showed acceptable storage variables with no differences between CompoFlow and control.
  • CompoFlow closures showed a highly reproducible aperture after opening.

Takeaway

A new blood bag closure makes it easier to open and doesn't harm the blood inside, helping to make blood collection faster and safer.

Methodology

Blood was collected and processed using CompoFlow and standard blood bag systems, with various quality assessments performed on RBCs and PLTs over time.

Potential Biases

Potential bias due to the involvement of the manufacturer in the study.

Limitations

The study did not test the automated opening device in its final form.

Participant Demographics

Blood donors were deferred for reasons other than their own health.

Statistical Information

P-Value