Evaluating the Fit and Scanning Time of CAD/CAM Ceramic Crowns
Author Information
Author(s): Pereira Leandro Maruki, de Melo Bárbara Inácio, Oliveira Alex Antônio Maciel, Mendonça Gustavo, Raposo Luís Henrique Araújo, Prudente Marcel Santana, das Neves Flávio Domingues
Primary Institution: Federal University of Uberlandia
Hypothesis
The scanning time and marginal fit of ceramic crowns would not differ between the intraoral scanning systems tested.
Conclusion
The Primescan system provided superior marginal fit and shorter scanning time compared to the Omnicam systems.
Supporting Evidence
- The Primescan system showed the best vertical fit values at 12.3 ± 6.6 μm.
- Scanning times were shortest for the Primescan at 27.8 ± 1.9 seconds.
- Significant differences in vertical fit were observed between the Primescan and both Omnicam systems.
Takeaway
This study looked at how well different dental scanning machines fit crowns and how long they take to scan. The Primescan was the best at both fitting and speed.
Methodology
The study involved scanning a standardized die with different intraoral scanning systems, designing and milling crowns, and measuring their fit using micro-computed tomography.
Limitations
The study was conducted in vitro, and clinical variables such as patient movements and saliva were not present.