Evaluation of Marginal Fit of CAD/CAM Ceramic Crowns and Scanning Time Using Different Intraoral Scanning Systems
2024

Evaluating the Fit and Scanning Time of CAD/CAM Ceramic Crowns

Sample size: 10 publication 10 minutes Evidence: high

Author Information

Author(s): Pereira Leandro Maruki, de Melo Bárbara Inácio, Oliveira Alex Antônio Maciel, Mendonça Gustavo, Raposo Luís Henrique Araújo, Prudente Marcel Santana, das Neves Flávio Domingues

Primary Institution: Federal University of Uberlandia

Hypothesis

The scanning time and marginal fit of ceramic crowns would not differ between the intraoral scanning systems tested.

Conclusion

The Primescan system provided superior marginal fit and shorter scanning time compared to the Omnicam systems.

Supporting Evidence

  • The Primescan system showed the best vertical fit values at 12.3 ± 6.6 μm.
  • Scanning times were shortest for the Primescan at 27.8 ± 1.9 seconds.
  • Significant differences in vertical fit were observed between the Primescan and both Omnicam systems.

Takeaway

This study looked at how well different dental scanning machines fit crowns and how long they take to scan. The Primescan was the best at both fitting and speed.

Methodology

The study involved scanning a standardized die with different intraoral scanning systems, designing and milling crowns, and measuring their fit using micro-computed tomography.

Limitations

The study was conducted in vitro, and clinical variables such as patient movements and saliva were not present.

Statistical Information

P-Value

p<0.001

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.3390/jfb15120359