Prevention of retinopathy of prematurity in preterm infants through changes in clinical practice and SpO2 technology
2011

Reducing Retinopathy of Prematurity in Preterm Infants with Pulse Oximetry

Sample size: 571 publication Evidence: moderate

Author Information

Author(s): Castillo Armando, Deulofeut Richard, Critz Ann, Sola Augusto

Primary Institution: Emory University

Hypothesis

The relative risk reduction in severe retinopathy of prematurity is associated with the saturation technology utilized in neonatal intensive care units.

Conclusion

A change in clinical practice and the use of pulse oximetry with Masimo SET technology significantly reduced severe ROP and the need for laser therapy in preterm infants.

Supporting Evidence

  • Severe ROP decreased from 12% to 5% at Centre B after implementing SET technology.
  • Need for laser treatment decreased from 5% to 3% at Centre B after the change to SET technology.
  • At Centre A, severe ROP decreased from 13% to 6% after changing to SET technology.

Takeaway

Using better pulse oximetry technology can help prevent serious eye problems in premature babies.

Methodology

Inborn infants <1250 g who had eye exams were compared at two centres over three periods with different SpO2 targets and pulse oximetry technologies.

Potential Biases

The study design may not definitively isolate the pulse oximetry technology as the only factor contributing to the decrease in ROP.

Limitations

This was not a randomized controlled trial, and different bedside nurses and respiratory therapists were involved at the two centres.

Participant Demographics

Inborn infants with birth weight <1250 g, with similar birth weight and gestational age across periods.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1111/j.1651-2227.2010.02001.x

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication