A SIMPLE METHOD FOR THE ACCURATE VOLUMETRIC MEASUREMENT OF SUPERFICIAL TUMOURS
1964

Accurate Measurement of Tumour Size

publication Evidence: moderate

Author Information

Author(s): E. J. Delorme

Primary Institution: Chester Beatty Research Institute, Institute of Cancer Research, Royal Cancer Hospital

Hypothesis

A new method for measuring the volume of superficial tumours can improve accuracy over existing techniques.

Conclusion

The new volumetric measurement method is inexpensive, time-saving, and accurate to about 0.2 ml.

Supporting Evidence

  • The new method allows for rapid measurements on a large number of animals with minimal disturbance.
  • Correlation is high between instrument values and actual volumes of plasticine models of tumours.

Takeaway

This study shows a new way to measure tumours that is quick and gives better results than older methods.

Methodology

A modified plethysmograph was used to measure tumour volume by comparing water levels in a pipette before and after applying the device to the tumour.

Limitations

The method may be difficult to use on small animals with thin abdominal walls.

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication