Accuracy of Non-Invasive Temperature Monitoring in Infants and Toddlers
Author Information
Author(s): Tasuku Fujii, Masashi Takakura, Tomoya Taniguchi, Kimitoshi Nishiwaki
Primary Institution: Nagoya University Hospital
Hypothesis
The Temple Touch Pro (TTP) accurately reflects esophageal temperature in infants and toddlers.
Conclusion
Core temperature measurements using TTP in infants and toddlers were more accurate with esophageal than with rectal measurements.
Supporting Evidence
- The bias between TTP and Tesoph was 0.09 ± 0.28 °C.
- 94.0% of measurements were within ±0.5 °C for TTP vs. Tesoph.
- TTP was more accurate with esophageal temperature than rectal temperature.
- The study included 40 patients with no exclusions.
- Infants showed a correlation coefficient of 0.919 between TTP and Tesoph.
- 68.5% of measurements were within ±0.5 °C for TTP vs. Trect.
- Adverse events related to temperature measurement were not observed.
- TTP allows continuous monitoring without invasiveness.
Takeaway
This study shows that a new way to measure body temperature without putting anything inside the body works well for young kids.
Methodology
This was a single-center prospective observational study comparing TTP with esophageal and rectal temperature measurements in pediatric patients.
Potential Biases
The esophageal temperature probe placement was not standardized, which could introduce variability.
Limitations
The study was conducted under general anesthesia, which may affect temperature regulation.
Participant Demographics
40 pediatric patients under 3 years old, including 20 infants and 20 toddlers.
Statistical Information
P-Value
0.05
Confidence Interval
95% limits of agreement were -0.48 to 0.65 °C for TTP vs. Tesoph.
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website