Accuracy of non-invasive core temperature monitoring in infant and toddler patients: a prospective observational study
2024

Accuracy of Non-Invasive Temperature Monitoring in Infants and Toddlers

Sample size: 40 publication 10 minutes Evidence: moderate

Author Information

Author(s): Tasuku Fujii, Masashi Takakura, Tomoya Taniguchi, Kimitoshi Nishiwaki

Primary Institution: Nagoya University Hospital

Hypothesis

The Temple Touch Pro (TTP) accurately reflects esophageal temperature in infants and toddlers.

Conclusion

Core temperature measurements using TTP in infants and toddlers were more accurate with esophageal than with rectal measurements.

Supporting Evidence

  • The bias between TTP and Tesoph was 0.09 ± 0.28 °C.
  • 94.0% of measurements were within ±0.5 °C for TTP vs. Tesoph.
  • TTP was more accurate with esophageal temperature than rectal temperature.
  • The study included 40 patients with no exclusions.
  • Infants showed a correlation coefficient of 0.919 between TTP and Tesoph.
  • 68.5% of measurements were within ±0.5 °C for TTP vs. Trect.
  • Adverse events related to temperature measurement were not observed.
  • TTP allows continuous monitoring without invasiveness.

Takeaway

This study shows that a new way to measure body temperature without putting anything inside the body works well for young kids.

Methodology

This was a single-center prospective observational study comparing TTP with esophageal and rectal temperature measurements in pediatric patients.

Potential Biases

The esophageal temperature probe placement was not standardized, which could introduce variability.

Limitations

The study was conducted under general anesthesia, which may affect temperature regulation.

Participant Demographics

40 pediatric patients under 3 years old, including 20 infants and 20 toddlers.

Statistical Information

P-Value

0.05

Confidence Interval

95% limits of agreement were -0.48 to 0.65 °C for TTP vs. Tesoph.

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1007/s00540-024-03404-7