An evaluation of the Kemtek 1000 sample processor
1988

Evaluation of the Kemtek 1000 Sample Processor

publication Evidence: moderate

Author Information

Author(s): M. J. Wheeler, Linzi Waiters

Primary Institution: Department of Chemical Pathology, St Thomas' Hospital

Hypothesis

The Kemtek 1000 Sample Processor will demonstrate high precision, accuracy, speed, and reliability in automated radioimmunoassays.

Conclusion

The Kemtek 1000 Sample Processor is reliable and precise, with minor issues that were quickly resolved.

Supporting Evidence

  • Precision was better than 1% at all volumes tested.
  • Accuracy was within ±5%.
  • A 100-tube assay could be set up within 15 minutes.
  • Carry-over could be reduced to <0.01% with sufficient wash steps.
  • Minor problems over 12 months were quickly resolved.

Takeaway

The Kemtek 1000 Sample Processor helps labs do tests faster and more accurately, making it easier for staff to manage their work.

Methodology

The processor was evaluated for precision, accuracy, speed, and reliability through various assay protocols.

Limitations

The processor is large, which may not be suitable for smaller labs, and there were some initial technical issues with the computer.