Evaluation of the Kemtek 1000 Sample Processor
Author Information
Author(s): M. J. Wheeler, Linzi Waiters
Primary Institution: Department of Chemical Pathology, St Thomas' Hospital
Hypothesis
The Kemtek 1000 Sample Processor will demonstrate high precision, accuracy, speed, and reliability in automated radioimmunoassays.
Conclusion
The Kemtek 1000 Sample Processor is reliable and precise, with minor issues that were quickly resolved.
Supporting Evidence
- Precision was better than 1% at all volumes tested.
- Accuracy was within ±5%.
- A 100-tube assay could be set up within 15 minutes.
- Carry-over could be reduced to <0.01% with sufficient wash steps.
- Minor problems over 12 months were quickly resolved.
Takeaway
The Kemtek 1000 Sample Processor helps labs do tests faster and more accurately, making it easier for staff to manage their work.
Methodology
The processor was evaluated for precision, accuracy, speed, and reliability through various assay protocols.
Limitations
The processor is large, which may not be suitable for smaller labs, and there were some initial technical issues with the computer.