The Influence of the Degree of Dental Implant Insertion Compression on Primary Stability Measured by Resonance Frequency and Progressive Insertion Torque: In Vitro Study
2024

Impact of Dental Implant Insertion Compression on Stability

Sample size: 120 publication Evidence: high

Author Information

Author(s): José Rosas-Díaz, Maria Eugenia Guerrero, Nancy Córdova-Limaylla, Maisely Galindo-Gómez, Marco García-Luna, César Cayo-Rojas, Oliver Schierz

Primary Institution: Universidad Privada San Juan Bautista

Hypothesis

There are no significant changes in the primary stability according to the insertion torque value and resonance frequency analysis of dental implants placed in standardized blocks of bone quality equivalent to type II-A bone when using three undersized surgical protocols.

Conclusion

Optimal surgical protocols for dental implant insertion can significantly enhance primary stability, achieving a 63.9% assurance of stability based on insertion torque and resonance frequency analysis.

Supporting Evidence

  • The average insertion torque was significantly higher with a 0.8 mm under-milling protocol.
  • The insertion torque was significantly lower with a 0.2 mm under-milling protocol.
  • The ISQ did not show significant differences across the different under-milling protocols.

Takeaway

This study shows that how you prepare the bone for dental implants can make them more stable, especially if you choose the right method based on the bone quality.

Methodology

The study used 120 dental implants of various sizes placed in polyurethane blocks to evaluate primary stability through insertion torque and resonance frequency analysis.

Limitations

The study was conducted in vitro, which may not fully replicate human bone conditions.

Statistical Information

P-Value

p < 0.001

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.3390/biomedicines12122878