Evaluating the Accuracy of Spirometers in Veterinary Hospitals
Author Information
Author(s): Raillard Mathieu, Mosing Martina, Raisis Anthea, Auckburally Adam, Borland Karla, Canfrán Susana, Downing Frances, García de Carellán Mateo Alejandra, MacFarlane Paul, McFadzean William, Merlin Tristan, Portier Karine, Robertson Josephine, Soares Joao Henrique Neves, Steblaj Barbara, Zoff Aurora, Levionnois Olivier L.
Primary Institution: Murdoch University
Hypothesis
Displayed tidal volumes would align with the performance claimed by the manufacturer (±6% or ±4–30 mL, whichever is larger).
Conclusion
The study found that the D-lite flow sensor performed better than the Pedi-lite sensor, particularly for volumes between 300 and 750 mL.
Supporting Evidence
- Only 37% of measurements with the Pedi-lite sensor met manufacturer specifications.
- 94% of measurements with the D-lite sensor were within manufacturer specifications.
- Significant differences were found between inspired and expired volumes for both sensors.
Takeaway
This study looked at how well spirometers measure air volumes in animals during anesthesia, finding that some devices work better than others.
Methodology
The performance of near-patient spirometers was assessed using calibration syringes across 67 monitors in 14 veterinary hospitals.
Potential Biases
High inter-monitor variability may affect the accuracy of spirometry readings.
Limitations
Variability in measurements could be influenced by different investigators and the calibration status of the syringes was not verified.
Participant Demographics
Monitors from 14 veterinary hospitals across six countries were tested.
Statistical Information
Statistical Significance
p<0.05