Screening for Low Bone Mass in a Community Without DXA
Author Information
Author(s): Yang Nan-Ping, Jen Ian, Chuang Shao-Yuan, Chen Shui-Hu, Chou Pesus
Primary Institution: National Yang-Ming University
Hypothesis
Can quantitative ultrasonography (QUS) effectively screen for low bone mass in a community lacking access to dual-energy X-ray absorptiometry (DXA)?
Conclusion
Age-related decreases in calcaneal ultrasonometry, which reflected the prevalence of low bone mass, were more obvious in women than in men.
Supporting Evidence
- The correlation coefficient between two QUS machines was 0.90.
- Calcaneal BUA was significantly correlated with BMD in the femoral neck (r = 0.67) and lumbar spine (r = 0.59).
- Annual losses of calcaneal BUA were 0.83% for women and 0.27% for men.
Takeaway
This study looked at how well a special ultrasound can check for weak bones in a place where the usual bone scan isn't available, finding that older women are more affected.
Methodology
A cross-sectional, community-based study where 6493 residents underwent ultrasonographic measurements and completed a questionnaire.
Potential Biases
Potential bias due to exclusion of military personnel and those with chronic disorders.
Limitations
The study could not calculate the true response rate due to the transient population in Kinmen and lacked normative data for DXA.
Participant Demographics
2792 men (43.0%) and 3701 women (57.0%) aged 40 years or older.