Scoliosis instrumentation alters primary and coupled motions of the spine: An in vitro study using entire thoracolumbar spine and rib cage specimens
2024

Effects of Scoliosis Instrumentation on Spine Motion

Sample size: 6 publication Evidence: moderate

Author Information

Author(s): Liebsch Christian, Obid Peter, Vogt Morten, Schlager Benedikt, Wilke Hans‐Joachim

Primary Institution: Institute of Orthopaedic Research and Biomechanics, Trauma Research Centre Ulm, Ulm University Medical Centre

Hypothesis

The study aims to quantify the effects of rigid posterior instrumentation on the three-dimensional post-operative spinal flexibility in adolescent idiopathic scoliosis.

Conclusion

Instrumentation reduces the primary flexibility and diminishes the natural coupling behavior between lateral bending and axial rotation, primarily in the upper thoracic spine.

Supporting Evidence

  • Primary ranges of motion were significantly reduced in all six motion directions in the thoracic spine for both instrumentations.
  • Coupled motions were significantly reduced, especially in the upper thoracic spine for T2-L1 instrumentation.
  • Instrumentation was found to significantly reduce primary flexibility in all motion directions.

Takeaway

When doctors use rods to fix a curved spine, it makes the spine less flexible and changes how it moves, which could cause problems later.

Methodology

The study tested six fresh frozen human spine specimens under different loading conditions to measure primary and coupled motions with and without instrumentation.

Potential Biases

The use of specimens without clinically relevant deformities may limit the generalizability of the findings to actual scoliosis cases.

Limitations

The study did not test scoliotic specimens and used pure moments that may not fully reflect physiological loading conditions.

Participant Demographics

Young adult donors aged 26–45 years, including two females and four males.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1002/jsp2.70028

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication