Comparing the old and new generation SELDI-TOF MS: implications for serum protein profiling
2008

Comparing Old and New SELDI-TOF MS for Serum Protein Profiling

Sample size: 179 publication 10 minutes Evidence: moderate

Author Information

Author(s): Marie-Christine Gast, Judith Engwegen, Jan HM Schellens, Jos H Beijnen

Primary Institution: The Netherlands Cancer Institute

Hypothesis

The new PCS 4000 SELDI-TOF MS instrument will outperform the old PBS-IIc in serum protein profiling.

Conclusion

The PCS 4000 instrument showed superior performance in peak detection after serum fractionation, but did not outperform the PBS-IIc in crude serum analysis.

Supporting Evidence

  • The PCS 4000 detected more peaks in serum fractions compared to the PBS-IIc.
  • Both instruments were tested on two different cancer patient cohorts.
  • The study found no significant difference in performance for crude serum analysis.

Takeaway

Scientists tested two machines that look at proteins in blood to see which one is better. The new machine is better at finding certain proteins when the blood is treated in a special way.

Methodology

Two sample sets were profiled using both SELDI-TOF MS instruments, and the performance was evaluated based on the number of peaks detected and their reproducibility.

Limitations

The study could not confirm the improved performance of the PCS 4000 over the PBS-IIc in crude serum analysis.

Participant Demographics

The study included 45 colorectal cancer patients and 43 matched controls, as well as 45 breast cancer patients and 46 matched normal women.

Statistical Information

P-Value

p<0.01

Statistical Significance

p<0.01

Digital Object Identifier (DOI)

10.1186/1755-8794-1-4