Reproducibility of quantitative myocardial perfusion and coronary flow capacity by positron emission tomography: 3D digital silicon photomultiplier solid state vs. legacy 2D analogue systems for clinical practice and trials
2024

Comparing 2D and 3D PET-CT for Heart Imaging

Sample size: 171 publication 10 minutes Evidence: moderate

Author Information

Author(s): Roby Amanda Weatherhead, Harmon Lindsey, Sander Kelly, Bui Linh, Kitkungvan Danai, Patel Monica, Narula Jagat, Johnson Nils P, Gould K Lance

Primary Institution: Weatherhead PET Center for Preventing and Reversing Atherosclerosis, McGovern Medical School, University of Texas Health Science Center at Houston

Hypothesis

The study aims to evaluate the accuracy and precision of myocardial perfusion measurements using 2D and 3D PET-CT scanners.

Conclusion

3D SiPM PET-CT is as reproducible as 2D PET-CT for measuring myocardial perfusion, aiding in clinical decision-making.

Supporting Evidence

  • 3D SiPM PET-CT showed no significant bias compared to 2D PET-CT.
  • Methodological variability was assessed with a coefficient of variation of ±14% for same-day tests.
  • 154 paired PET scans were analyzed for different-day variability with a coefficient of variation of ±20%.
  • Coronary flow reserve measurements showed a small bias but were statistically significant.

Takeaway

This study shows that new 3D heart imaging technology works just as well as older 2D technology for checking heart health.

Methodology

Participants underwent paired serial rest or stress PET perfusion imaging using rubidium-82 on both 2D and 3D PET-CT scanners.

Potential Biases

Potential bias due to the study being conducted at a single center with a specific imaging protocol.

Limitations

The study was conducted at a single institution and may not be generalizable to other settings.

Participant Demographics

{"male_percentage":64,"average_age":58,"average_BMI":29}

Statistical Information

P-Value

0.041

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1093/ehjimp/qyae115