New Westgard Rules for Immunological Parameters
Author Information
Author(s): Cristelli Lisa, Occhipinti Francesca, Tumiatti Daniel, Antonia De Luisi, Jani Erika, Daves Massimo
Primary Institution: Clinical Biochemistry Laboratory, Provincial Hospital of Bolzano (SABES-ASDAA), Bolzano, Italy
Hypothesis
Are the Westgard rules proposed by Bio-Rad’s Westgard Advisor software more efficient in monitoring analytical performance than those previously in use?
Conclusion
The newly suggested rejection rules did not lead to an improvement in monitoring analytical performance.
Supporting Evidence
- The study assessed new Westgard rules against those previously in use.
- Analytical quality was not significantly improved with the new rules.
- Improvements in statistical values were inhomogeneous across parameters.
Takeaway
The study tested new rules for checking lab results but found they didn't make things better.
Methodology
The study evaluated five immunological parameters using different Westgard rules over a two-month period.
Potential Biases
The study faced challenges in adhering to the new rejection rules due to their complexity.
Limitations
Different lots of reagents were used during the evaluation period, which could affect the results.
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website