Wide FOV metalens for near-infrared capsule endoscopy: advancing compact medical imaging
2024

Wide FOV Metalens for Capsule Endoscopy

publication Evidence: high

Author Information

Author(s): Moghaddasi Mojtaba, Perez Coca Erik Edilson, Ye Danni, Flores Diego Alejandro, Wu Xudong, Jalal Abdul, Ren Ziming, Abrinaei Fahimeh, Hu Bin

Primary Institution: Beijing Institute of Technology, China

Hypothesis

Can a wide field-of-view metalens improve the performance of capsule endoscopy?

Conclusion

The study demonstrates that a metalens can achieve a 165° field of view and improved imaging quality for capsule endoscopy.

Supporting Evidence

  • The metalens achieved a 165° FOV with a high modulation transfer function of 300 lines per millimeter.
  • The performance of the metalens-based system was evaluated using two bandwidths, showing MTF values of 0.2 and 0.3 at 250 lp/mm.
  • Compared to traditional endoscopes, the nano-optic capsule endoscope demonstrated superior performance in terms of FOV, contrast, and resolution.

Takeaway

This study shows that a new type of lens can help doctors see better inside the body using a tiny camera.

Methodology

The study involved the design, fabrication, and characterization of a metalens optimized for capsule endoscopy, using narrow-bandpass imaging.

Limitations

The alignment of the metalens with the CMOS sensor posed challenges, and image quality may be affected by artifacts.

Digital Object Identifier (DOI)

10.1515/nanoph-2024-0393

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication