Comparing Two Methods for Measuring the Visual Axis
Author Information
Author(s): Rodrigues Pablo Felipe, Moscovici Bernardo Kaplan, Lamazales Luciano, Freitas Marcela Mara Silva, Gomes José Álvaro Pereira, Nosé Walton, Campos Mauro Silveira
Primary Institution: Escola Paulista de Medicina, Universidade Federal de São Paulo
Hypothesis
To compare the differences between the apparent and actual chord µ.
Conclusion
Both measurement devices can be used in daily practice despite their considerable differences.
Supporting Evidence
- The mean actual chord µ was 226.21 ± 128.53 µm and the apparent chord µ was 278.66 ± 123.90 µm.
- Pearson’s correlation coefficient of 0.647 indicates a moderate positive linear relationship between the two measurement methods.
- The mean difference between the two methods was 52.45 µm.
Takeaway
The study looked at two ways to measure a part of the eye and found that while they give different results, both can be useful.
Methodology
This was a prospective, comparative, non-randomized, and non-interventional study using imaging examinations with Pentacam and HD Analyzer.
Limitations
The study excluded patients with certain eye conditions and those using contact lenses, which may limit the generalizability of the findings.
Participant Demographics
The study included 58 patients aged 21-71 years, with an equal number of males and females.
Statistical Information
P-Value
0.01
Confidence Interval
260.55, -155.64
Statistical Significance
p=0.01
Digital Object Identifier (DOI)