Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer
2005

Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer

Sample size: 45 publication Evidence: high

Author Information

Author(s): Yutaka Nagai, Hiroshi Kondo, Noriyuki Tatsumi

Primary Institution: Nihon Kohden Corporation

Hypothesis

The Celltac F hematology analyzer provides accurate platelet counts compared to the ICSH/ISLH reference method.

Conclusion

The Celltac F hematology analyzer is efficient and provides accurate platelet counts, making it suitable for clinical laboratory use.

Supporting Evidence

  • The correlation coefficient for the Celltac F versus the ICSH/ISLH reference method was 0.99.
  • The correlation coefficient for the Celltac F versus the manual visual method was 0.93.
  • Platelet counts by the ICSH/ISLH reference method showed acceptable stability for 6 hours.

Takeaway

The Celltac F machine counts platelets accurately, which helps doctors know how many platelets are in your blood.

Methodology

Platelet counts were compared between the Celltac F analyzer and the ICSH/ISLH reference method using blood samples from healthy volunteers.

Limitations

The study may not be applicable for routine tests due to the cost and time-consuming sample preparation process.

Participant Demographics

Healthy adult volunteers were used for blood samples.

Statistical Information

Statistical Significance

p<0.05