Normal Values for Macular Perimetry Using MP-3 Microperimeter
Author Information
Author(s): Rodrigues Neto Taurino dos Santos, da Silva Neto Epitácio Dias, Higashi Alex Haruo, Megnis Bianca Partezani, Haddad Maria Aparecida Onuki, Monteiro Mário Luiz Ribeiro, Zacharias Leandro Cabral
Primary Institution: Faculdade de Medicina, Universidade de São Paulo, São Paulo, SP, Brazil
Hypothesis
This study aimed to determine values for light sensitivity thresholds and fixation stability using the MP-3 in healthy individuals.
Conclusion
The results of this study provide a normal and age-matched database of MP-3 microperimetry.
Supporting Evidence
- The global mean sensitivity was 29.01 ± 1.44 dB.
- The mean central sensitivity at 2° was 28.5 ± 1.77 dB in the right eye and 28.75 ± 1.98 dB in the left eye.
- The total median fixation stability values were 80% and 96% for 2° and 4°, respectively.
- Linear regression analysis revealed an age-related global sensitivity decline per year of -0.051 dB ± 0.018 (OD) and -0.078 dB ± 0.021 (OS).
Takeaway
The study measured how well healthy people's eyes respond to light using a special device, helping doctors understand normal eye function.
Methodology
Thirty-seven healthy volunteers underwent full-threshold microperimetry using a 4-2 staircase strategy with the standard Goldmann III stimulus size and 68 test points.
Limitations
The relatively small number of participants and having only one measurement per eye.
Participant Demographics
Healthy adults aged 28-68 years.
Statistical Information
P-Value
<0.05
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website