Evaluation of pulse-oximetry oxygen saturation taken through skin protective covering
2006

Evaluating Pulse-Oximetry Accuracy with Skin Protection

Sample size: 50 publication Evidence: moderate

Author Information

Author(s): James Jyotsna, Tiwari Lokesh, Upadhyay Pramod, Sreenivas Vishnubhatla, Bhambhani Vikas, Puliyel Jacob M

Primary Institution: St Stephen's Hospital Tis Hazari, Delhi, India

Hypothesis

Does using a protective covering affect the accuracy of pulse-oximetry readings in neonates?

Conclusion

Using Micropore or gauze does not significantly affect the accuracy of pulse-oximetry readings in neonates.

Supporting Evidence

  • The mean difference in readings with Micropore was 0.06 and with gauze was 0.04 compared to the standard method.
  • Bland-Altman plots indicated no significant loss of accuracy with protective coverings.
  • The differences observed were less than the variability seen in repeatability testing of the standard method.

Takeaway

This study shows that putting tape or gauze on a baby's skin before taking oxygen readings doesn't change the results much.

Methodology

Fifty full-term neonates had pulse-oximetry readings taken directly and through protective coverings, with comparisons made using Bland-Altman plots.

Limitations

The study did not use blood gas readings as a gold standard for comparison.

Participant Demographics

All participants were full-term neonates less than a week old.

Digital Object Identifier (DOI)

10.1186/1471-2431-6-14