Electrochemical Disinfection of Dental Implants – a Proof of Concept
2011

Electrochemical Disinfection of Dental Implants

Sample size: 9 publication Evidence: moderate

Author Information

Author(s): Mohn Dirk, Zehnder Matthias, Stark Wendelin J., Imfeld Thomas

Primary Institution: Institute for Chemical and Bioengineering, ETH Zurich

Hypothesis

Electrolysis can reduce viable counts of adhering bacteria, and this reduction should be greater if active oxidative species are generated.

Conclusion

This laboratory study shows that electrochemical treatment may provide access to a new way to decontaminate dental implants in situ.

Supporting Evidence

  • Electrochemical treatment reduced viable E. coli counts by up to 99%.
  • At the anode, a current of 7.5 mA or higher resulted in complete bacterial kill.
  • The alkaline environment at the cathode also contributed to bacterial reduction.

Takeaway

The study tested a new way to clean dental implants using electricity, which can kill bacteria more effectively than traditional methods.

Methodology

Dental titanium implants were coated with a biofilm of E. coli and treated with electric currents of varying strengths to assess bacterial reduction.

Limitations

The study used a single strain of E. coli, which may not fully represent the mixed microbial environment found in clinical situations.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1371/journal.pone.0016157

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication