Intra-Examiner and Inter-Examiner Reproducibility of Paraspinal Thermography
2011

Reproducibility of Paraspinal Thermography

Sample size: 93 publication Evidence: high

Author Information

Author(s): Matthew McCoy, Ismay Campbell, Pamela Stone, Curtis Fedorchuk, Sameera Wijayawardana, Kirk Easley

Primary Institution: Life University College of Chiropractic

Hypothesis

The study aims to evaluate the intra-examiner and inter-examiner reproducibility of paraspinal thermography using an infrared scanner.

Conclusion

The study revealed excellent intra-examiner and inter-examiner reproducibility of paraspinal thermography using a commercially available unit.

Supporting Evidence

  • The intra class correlation coefficients (ICCs) for agreement ranged from 0.959 to 0.976.
  • Concordance correlation coefficients (CCCs) ranged from 0.783 to 0.859.
  • The average bilateral paraspinal skin temperature was found to be 89.78° F.

Takeaway

This study shows that two doctors can get the same results when using a special thermometer on people's backs, which means it's a reliable way to check for problems.

Methodology

Two clinicians conducted thermal scans on 100 subjects using an infrared scanner, and data were analyzed for reproducibility.

Potential Biases

Results may be biased due to the long-term use of the equipment by the examiners.

Limitations

The study did not address clinical applications or the sensitivity and specificity of thermal scanning for diagnosing spinal disorders.

Participant Demographics

93 subjects, including 57 males and 36 females, aged 21 to 60 years with a mean age of 32.

Statistical Information

Confidence Interval

0.783 to 0.859

Digital Object Identifier (DOI)

10.1371/journal.pone.0016535