Spinal instrumentation length affects adjacent segment range of motion and intradiscal pressure
2024

How Spinal Instrumentation Length Affects Spine Movement and Pressure

Sample size: 6 publication Evidence: moderate

Author Information

Author(s): Liebsch Christian, Obid Peter, Vogt Morten, Schlager Benedikt, Wilke Hans-Joachim

Primary Institution: Institute of Orthopaedic Research and Biomechanics, Ulm University Medical Centre

Hypothesis

Increasing spinal instrumentation length affects adjacent segment mobility and intervertebral disc loading.

Conclusion

Longer spinal instrumentation leads to reduced overall spinal flexibility and increased pressure in adjacent segments.

Supporting Evidence

  • All tested posterior instrumentation lengths caused significant reductions of the overall range of motion compared to the condition without posterior fixation.
  • Significant increases in adjacent segment range of motion were found for longer instrumentation lengths.
  • Instrumented levels exhibited significantly increased intradiscal pressure compared to non-instrumented levels.

Takeaway

When doctors put in rods to fix the spine, making them longer can make it harder for the spine to move and put more pressure on nearby parts.

Methodology

The study tested six fresh frozen human spine specimens under controlled loading conditions to measure range of motion and intradiscal pressure.

Limitations

The sample size was small, and the specimens did not have clinically relevant spinal deformities.

Participant Demographics

Young adult donors aged 26-45 years, with 2 females and 4 males.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1038/s41598-024-82132-0

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication