Reliability of Ankle-Brachial Index Measurements
Author Information
Author(s): Beth D Weatherley, Lloyd E Chambless, Gerardo Heiss, Diane J Catellier, Curtis R Ellison
Primary Institution: Duke Clinical Research Institute, Duke University Medical Center
Hypothesis
The reliability coefficient for the ankle-brachial index (ABI) computed from DINAMAP readings is not known.
Conclusion
The reliability estimates suggest the need for repeated measures of the ABI in clinical practice before diagnosing peripheral artery disease.
Supporting Evidence
- The estimated reliability coefficient for the ABI was 0.612.
- The reliability for the ABI computed as the ratio of the average of two ankle SBPs to two arm SBPs was estimated as 0.70.
- The study suggests that repeated measures of the ABI are necessary for accurate diagnosis.
Takeaway
Doctors should check the ankle-brachial index (ABI) multiple times to make sure it's accurate before deciding on treatment for blood flow problems.
Methodology
Repeat ABIs were taken within 1 year using a common protocol and automated oscillometric devices.
Potential Biases
Potential bias due to measurement error and variability in technician performance.
Limitations
The study had a limited number of participants in certain subgroups, which restricted subgroup analysis.
Participant Demographics
58.8% women, 84.0% white, mean age 61.3 years.
Statistical Information
P-Value
0.02
Confidence Interval
95% CI: 0.505, 0.699
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website