Evaluation of the UniCel DxH 800 Hematology Analyzer
Author Information
Author(s): Hedley B D Keeney, M Chin-Yee, I Brown, W Brown
Primary Institution: London Health Sciences Centre
Hypothesis
The study aims to evaluate the performance of the UniCel DxH 800 in accurately counting white blood cells, nucleated red blood cells, and platelets compared to existing analyzers.
Conclusion
The UniCel DxH 800 demonstrates significant improvements in accuracy and efficiency for blood cell enumeration, particularly in the presence of interference.
Supporting Evidence
- The DxH 800 showed high correlation with both the LH 780 and FCM for WBC counts.
- Flagging rates for NRBCs were significantly reduced with the DxH 800 compared to the LH 780.
- The DxH 800 demonstrated improved sensitivity and specificity for PLTs and NRBCs.
- Automated correction of RBC parameters was accurate in cases of extreme leukocytosis.
- The study included a diverse sample set from an 800-bed tertiary care hospital.
Takeaway
The UniCel DxH 800 is a new machine that counts blood cells more accurately and quickly, even when there are problems with the samples.
Methodology
The study involved analyzing 272 blood samples using the DxH 800 and comparing results with a predicate analyzer and flow cytometry.
Potential Biases
Potential biases in the comparison of results between the DxH 800 and predicate analyzers were noted, particularly in high platelet counts.
Limitations
The study did not compare WBC differentials, which may be addressed in future research.
Participant Demographics
The samples included both morphologically normal and hematologically abnormal patients.
Statistical Information
P-Value
0.918
Statistical Significance
p<0.05