Changeovers of vasoactive drug infusion pumps: impact of a quality improvement program
2008

Impact of Standardizing Vasoactive Drug Infusion Pump Changeovers

Sample size: 43 publication Evidence: moderate

Author Information

Author(s): Argaud Laurent, Cour Martin, Martin Olivier, Saint-Denis Marc, Ferry Tristan, Goyatton Agnes, Robert Dominique

Primary Institution: Hospices Civils de Lyon

Hypothesis

The variation in procedures for changing vasoactive infusion pumps could generate morbidity.

Conclusion

Standardizing the changeover of vasoactive infusion pumps significantly reduced the frequency of adverse incidents in the ICU.

Supporting Evidence

  • The frequency of incidents was reduced from 17.8% in phase 1 to 5.9% in phase 2.
  • No documented arrhythmias were observed during the study.
  • The study included a total of 913 CVIP events across two phases.

Takeaway

Changing medicine pumps in a specific way can help keep patients safer and prevent problems with their heart and blood pressure.

Methodology

A prospective before-and-after intervention study was conducted with adult patients receiving continuous infusion of vasoactive drugs, comparing incidents before and after implementing a standardized changeover method.

Potential Biases

Potential bias due to the observational nature of the study and the lack of randomization.

Limitations

The small number of patients studied may limit the generalizability of the findings.

Participant Demographics

Majority of patients had septic shock; 43 patients were included in total.

Statistical Information

P-Value

<0.0001

Statistical Significance

p<0.0001

Digital Object Identifier (DOI)

10.1186/cc6209