Comparing Blood Culture Systems for Detecting Infections
Author Information
Author(s): Dreyer Andries W, Ismail Nazir A, Nkosi Deliwe, Lindeque Kathy, Matthews Marliza, van Zyl Danie G, Hoosen Anwar A
Primary Institution: University of Pretoria and Tshwane Academic Division of the National Health Laboratory Service
Hypothesis
How does the VersaTREK blood culture system compare to the Bactec9240 system in recovering bloodstream pathogens?
Conclusion
The VersaTREK system is a useful alternative to the Bactec9240 for recovering bloodstream pathogens, despite a higher false positivity rate.
Supporting Evidence
- The VersaTREK system had 58 positive signals, with 14 being false positives.
- The Bactec9240 system yielded 43 positive signals with no false positives.
- The culture positivity rate for both systems was similar, at 25%.
Takeaway
This study looked at two systems that help find germs in blood. One system found more germs but also made more mistakes.
Methodology
Blood from patients suspected of sepsis was tested using both blood culture systems, and results were compared.
Potential Biases
Potential bias due to the study being laboratory-based without clinical data.
Limitations
The study could not assess false positives due to lack of prior antimicrobial usage data and had a small number of positive isolates.
Participant Demographics
Adult patients from medical wards and intensive care units, primarily in a high HIV prevalence setting.
Statistical Information
P-Value
0.45
Confidence Interval
-5.8583 - 12.9551
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website