Effect of zooming on texture features of ultrasonic images
2006

Impact of Zooming on Ultrasound Image Texture Analysis

Sample size: 18 publication Evidence: moderate

Author Information

Author(s): Kakkos Stavros K, Nicolaides Andrew N, Kyriacou Efthyvoulos, Pattichis Constantinos S, Geroulakos George

Primary Institution: Imperial College, London, UK

Hypothesis

Does image zooming during ultrasound scanning affect the texture features of carotid plaques?

Conclusion

Texture analysis of ultrasonic plaques should be performed under standardised resolution settings; otherwise, a resolution normalisation algorithm should be applied.

Supporting Evidence

  • Most second order statistics were significantly sensitive to image zooming during scanning.
  • Image interpolation restored changes, with bicubic interpolation being superior to bilinear interpolation.
  • Median change of features sensitive to zooming was 18.14%.

Takeaway

When taking pictures of carotid plaques with ultrasound, zooming can change how we see their texture, so we need to be careful about how we do it.

Methodology

Eighteen B-mode images of carotid plaques were zoomed and analyzed using bilinear and bicubic interpolation to assess texture features.

Limitations

The study had a small sample size and focused only on a specific zoom factor.

Participant Demographics

Asymptomatic patients with carotid plaques producing stenosis greater than 50%.

Statistical Information

P-Value

p < 0.05

Confidence Interval

18.14% (4.94–28.43)

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1476-7120-4-8