Impact of Zooming on Ultrasound Image Texture Analysis
Author Information
Author(s): Kakkos Stavros K, Nicolaides Andrew N, Kyriacou Efthyvoulos, Pattichis Constantinos S, Geroulakos George
Primary Institution: Imperial College, London, UK
Hypothesis
Does image zooming during ultrasound scanning affect the texture features of carotid plaques?
Conclusion
Texture analysis of ultrasonic plaques should be performed under standardised resolution settings; otherwise, a resolution normalisation algorithm should be applied.
Supporting Evidence
- Most second order statistics were significantly sensitive to image zooming during scanning.
- Image interpolation restored changes, with bicubic interpolation being superior to bilinear interpolation.
- Median change of features sensitive to zooming was 18.14%.
Takeaway
When taking pictures of carotid plaques with ultrasound, zooming can change how we see their texture, so we need to be careful about how we do it.
Methodology
Eighteen B-mode images of carotid plaques were zoomed and analyzed using bilinear and bicubic interpolation to assess texture features.
Limitations
The study had a small sample size and focused only on a specific zoom factor.
Participant Demographics
Asymptomatic patients with carotid plaques producing stenosis greater than 50%.
Statistical Information
P-Value
p < 0.05
Confidence Interval
18.14% (4.94–28.43)
Statistical Significance
p<0.05