New Westgard Rules for Immunological Parameters
Author Information
Author(s): Cristelli Lisa, Occhipinti Francesca, Tumiatti Daniel, Antonia De Luisi, Jani Erika, Daves Massimo
Primary Institution: Clinical Biochemistry Laboratory, Provincial Hospital of Bolzano (SABES-ASDAA), Bolzano, Italy
Hypothesis
The Westgard Advisor software creates new specific Westgard rules for five selected immunological parameters that guarantee the highest possible quality of the tests.
Conclusion
By using newly suggested rejection rules, we did not observe an improvement in monitoring of analytical performance.
Supporting Evidence
- The study did not show significant improvement in analytical performance monitoring.
- Improvements in statistical values were inhomogeneous across the parameters.
- Laboratory technicians became more attentive to internal quality control due to the study.
Takeaway
The study tested new rules for checking lab results but found that they didn't make things better.
Methodology
The study evaluated five immunological parameters using different Westgard rules over a two-month period.
Potential Biases
The study faced challenges in adhering to the new rejection rules due to their complexity.
Limitations
Different lots of reagents were used during the evaluation period, which could partly be responsible for the observed results.