Implementation of new Westgard rules suggested by the Westgard Advisor software for five immunological parameters
2023

New Westgard Rules for Immunological Parameters

publication Evidence: low

Author Information

Author(s): Cristelli Lisa, Occhipinti Francesca, Tumiatti Daniel, Antonia De Luisi, Jani Erika, Daves Massimo

Primary Institution: Clinical Biochemistry Laboratory, Provincial Hospital of Bolzano (SABES-ASDAA), Bolzano, Italy

Hypothesis

The Westgard Advisor software creates new specific Westgard rules for five selected immunological parameters that guarantee the highest possible quality of the tests.

Conclusion

By using newly suggested rejection rules, we did not observe an improvement in monitoring of analytical performance.

Supporting Evidence

  • The study did not show significant improvement in analytical performance monitoring.
  • Improvements in statistical values were inhomogeneous across the parameters.
  • Laboratory technicians became more attentive to internal quality control due to the study.

Takeaway

The study tested new rules for checking lab results but found that they didn't make things better.

Methodology

The study evaluated five immunological parameters using different Westgard rules over a two-month period.

Potential Biases

The study faced challenges in adhering to the new rejection rules due to their complexity.

Limitations

Different lots of reagents were used during the evaluation period, which could partly be responsible for the observed results.