Assessment of whole genome amplification-induced bias through high-throughput, massively parallel whole genome sequencing
2006

Assessing Bias in Whole Genome Amplification Techniques

publication Evidence: high

Author Information

Author(s): Pinard Robert, de Winter Alex, Sarkis Gary J, Gerstein Mark B, Tartaro Karrie R, Plant Ramona N, Egholm Michael, Rothberg Jonathan M, Leamon John H

Primary Institution: 454 Life Sciences, Branford CT, USA

Hypothesis

How do different whole genome amplification methods affect the accuracy and yield of DNA sequencing?

Conclusion

Multiple displacement amplification methods produced the least bias and highest yields compared to PCR-based methods.

Supporting Evidence

  • GenomiPhi generated 16.1 μg of DNA, a 640-fold amplification.
  • Repli-G amplified input DNA 2100 fold to 53.6 μg.
  • PEP generated 3.0 μg, a 120-fold increase.
  • DOP amplified the input DNA 92-fold to 2.3 μg.
  • All amplification methodologies induced statistically significant bias relative to the unamplified control.

Takeaway

This study looked at different ways to make more DNA from tiny samples and found that some methods work better than others without messing up the DNA too much.

Methodology

Whole genome sequencing was used to compare the bias introduced by different amplification methods on two bacterial genomes.

Potential Biases

PCR-based methods showed higher bias compared to multiple displacement amplification methods.

Limitations

The study focused only on two bacterial genomes and may not generalize to other organisms.

Statistical Information

P-Value

< 0.001

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.1186/1471-2164-7-216

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication