Retinal specific measurement of dark-adapted visual function: validation of a modified microperimeter
2011

Measuring Dark-Adapted Visual Function with a Modified Microperimeter

Sample size: 10 publication 10 minutes Evidence: moderate

Author Information

Author(s): Crossland Michael D, Luong Vy A, Rubin Gary S, Fitzke Fred W

Primary Institution: UCL Institute of Ophthalmology, London, UK

Hypothesis

Can a modified MP-1 microperimeter effectively measure scotopic visual function compared to a modified Humphrey field analyser?

Conclusion

The modified MP-1 microperimeter can measure dark-adapted visual function and provides results comparable to the modified Humphrey field analyser.

Supporting Evidence

  • The MP-1 microperimeter showed a central scotopic scotoma in all subjects.
  • Mean test time was significantly shorter for the MP-1 compared to the HFA.
  • The depth of the scotoma measured with the MP-1 was less than that measured with the HFA.

Takeaway

Researchers created a new way to test how well people see in the dark using a special device, and it works pretty well compared to an older method.

Methodology

Ten subjects underwent scotopic perimetry using a modified MP-1 microperimeter and a modified Humphrey field analyser in a counterbalanced order.

Potential Biases

Potential bias due to small sample size and recruitment from authors' acquaintances.

Limitations

The MP-1 microperimeter has a lower dynamic range, which may limit its ability to measure scotopic function in patients with eye disease.

Participant Demographics

Five younger subjects (<35 years) and five older subjects (>65 years) with no eye disease.

Statistical Information

P-Value

p < 0.0001

Statistical Significance

p < 0.01

Digital Object Identifier (DOI)

10.1186/1471-2415-11-5